T1 - Accuracy of a blood glucose monitoring system at hypoglycaemic glucose levels: analysis of seven clinical studies
AU - Cabezudo, J. Diago
AU - Bellary, S.
AU - Cameron, H.
AU - Ellison, J.
AU - Raja, P.
N1 - Abstracts of the 48th EASD Annual Meeting of the European Association for the Study of Diabetes
PY - 2012/10
Y1 - 2012/10
N2 - Background and aims: For individuals with diabetes detection of hypoglycemia
is crucial, and blood glucose monitoring systems (BGMSs) must
therefore be accurate in hypoglycemic ranges. Collecting sufficient data at
low blood glucose (BG) levels is difficult; which may explain why evidence of
BGMS performance during hypoglycemia is lacking. This analysis evaluated
the accuracy of 5 performance-equivalent BGMSs using OneTouch® Verio®
test strips at BG levels of < 3.9 mmol/L (70 mg/dL) in 7 separate clinical
studies.
Materials and methods: The studies were conducted between June 2009
and June 2011. BG testing was performed by trained staff in 2 clinical centers
in the UK. All studies were conducted in accordance with standard ISO
15197:2003 testing guidelines. Each study included 100 participants with diabetes.
For each participant, duplicate BGMS tests were performed using 3-4
test strip lots per study. BG reference values were obtained using a YSI 2300
STAT Glucose Analyzer before and after BGMS testing.
Results: Of a total of 5,400 measurements, 674 (12.5%) were at a BG level
of < 3.9 mmol/L (70 mg/dL). The number and percentage of results within
± 0.83 mmol/L (15 mg/dL) and ± 0.56 mmol/L (10 mg/dL) were calculated
at BG levels < 3.9 mmol/L (70 mg/dL), < 3.3 mmol/L (60 mg/dL), and < 2.8
mmol/L (50 mg/dL). Accuracy of the OneTouch Verio test strips at hypoglycemic
blood glucose levels are shown in the Table.
Conclusion: Compared with analysis based on a single study, using data from
7 studies, 5 BGMSs, and 15 different test strip lots provides a robust depiction
of test strip accuracy at BG levels < 3.9 mmol/L (70 mg/dL). The results indicate
that OneTouch Verio test strips are highly accurate in the hypoglycemic
range. Table. OneTouch Verio accuracy at BG levels < 3.9 mmol/L (70 mg/
dL).
AB - Background and aims: For individuals with diabetes detection of hypoglycemia
is crucial, and blood glucose monitoring systems (BGMSs) must
therefore be accurate in hypoglycemic ranges. Collecting sufficient data at
low blood glucose (BG) levels is difficult; which may explain why evidence of
BGMS performance during hypoglycemia is lacking. This analysis evaluated
the accuracy of 5 performance-equivalent BGMSs using OneTouch® Verio®
test strips at BG levels of < 3.9 mmol/L (70 mg/dL) in 7 separate clinical
studies.
Materials and methods: The studies were conducted between June 2009
and June 2011. BG testing was performed by trained staff in 2 clinical centers
in the UK. All studies were conducted in accordance with standard ISO
15197:2003 testing guidelines. Each study included 100 participants with diabetes.
For each participant, duplicate BGMS tests were performed using 3-4
test strip lots per study. BG reference values were obtained using a YSI 2300
STAT Glucose Analyzer before and after BGMS testing.
Results: Of a total of 5,400 measurements, 674 (12.5%) were at a BG level
of < 3.9 mmol/L (70 mg/dL). The number and percentage of results within
± 0.83 mmol/L (15 mg/dL) and ± 0.56 mmol/L (10 mg/dL) were calculated
at BG levels < 3.9 mmol/L (70 mg/dL), < 3.3 mmol/L (60 mg/dL), and < 2.8
mmol/L (50 mg/dL). Accuracy of the OneTouch Verio test strips at hypoglycemic
blood glucose levels are shown in the Table.
Conclusion: Compared with analysis based on a single study, using data from
7 studies, 5 BGMSs, and 15 different test strip lots provides a robust depiction
of test strip accuracy at BG levels < 3.9 mmol/L (70 mg/dL). The results indicate
that OneTouch Verio test strips are highly accurate in the hypoglycemic
range. Table. OneTouch Verio accuracy at BG levels < 3.9 mmol/L (70 mg/
dL).
UR - http://link.springer.com/article/10.1007%2Fs00125-012-2688-9
U2 - 10.1007/s00125-012-2688-9
DO - 10.1007/s00125-012-2688-9
M3 - Conference abstract
SN - 0012-186X
VL - 55
SP - S426
JO - Diabetologia
JF - Diabetologia
IS - Suppl.1
M1 - 1038
T2 - 48th EASD Annual Meeting
Y2 - 1 October 2012 through 5 October 2012
ER -