T1 - Wavefront analyzers induce instrument myopia
AU - Cerviño, Alejandro
AU - Hosking, Sarah L.
AU - Rai, Gurjeet K.
AU - Naroo, Shezhad A.
AU - Gilmartin, Bernard
PY - 2006/10
Y1 - 2006/10
N2 - PURPOSE: To assess the accuracy of three wavefront analyzers versus a validated binocular open-view autorefractor in determining refractive error in non-cycloplegic eyes.
METHODS: Eighty eyes were examined using the SRW-5000 open-view infrared autorefractor and, in randomized sequence, three wavefront analyzers: 1) OPD-Scan (NIDEK, Gamagori, Japan), 2) WASCA (Zeiss/Meditec, Jena, Germany), and 3) Allegretto (WaveLight Laser Technologies AG, Erlangen, Germany). Subjects were healthy adults (19 men and 21 women; mean age: 20.8 +/- 2.5 years). Refractive errors ranged from +1.5 to -9.75 diopters (D) (mean: +1.83 +/- 2.74 D) with up to 1.75 D cylinder (mean: 0.58 +/- 0.53 D). Three readings were collected per instrument by one examiner without anticholinergic agents. Refraction values were decomposed into vector components for analysis, resulting in mean spherical equivalent refraction (M) and J0 and J45 being vectors of cylindrical power at 0 degrees and 45 degrees, respectively.
RESULTS: Positive correlation was observed between wavefront analyzers and the SRW-5000 for spherical equivalent refraction (OPD-Scan, r=0.959, P<.001; WASCA, r=0.981, P<.001; Allegretto, r=0.942, P<.001). Mean differences and limits of agreement showed more negative spherical equivalent refraction with wavefront analyzers (OPD-Scan, 0.406 +/- 0.768 D [range: 0.235 to 0.580 D] [P<.001]; WASCA, 0.511 +/- 0.550 D [range: 0.390 to 0.634 D] [P<.001]; and Allegretto, 0.434 +/- 0.904 D [range: 0.233 to 0.635 D] [P<.001]). A second analysis eliminating outliers showed the same trend but lower differences: OPD-Scan (n=75), 0.24 +/- 0.41 D (range: 0.15 to 0.34 D) (P<.001); WASCA (n=78), 0.46 +/- 0.47 D (range: 0.36 to 0.57 D) (P<.001); and Allegretto (n=77), 0.30 +/- 0.62 D (range: 0.16 to 0.44 D) (P<.001). No statistically significant differences were noted for J0 and J45.
CONCLUSIONS: Wavefront analyzer refraction resulted in 0.30 D more myopia compared to SRW-5000 refraction in eyes without cycloplegia. This is the result of the accommodation excess attributable to instrument myopia. For the relatively low degrees of astigmatism in this study (<2.0 D), good agreement was noted between wavefront analyzers and the SRW-5000.
Copyright (C) 2006 SLACK Incorporated
AB - PURPOSE: To assess the accuracy of three wavefront analyzers versus a validated binocular open-view autorefractor in determining refractive error in non-cycloplegic eyes.
METHODS: Eighty eyes were examined using the SRW-5000 open-view infrared autorefractor and, in randomized sequence, three wavefront analyzers: 1) OPD-Scan (NIDEK, Gamagori, Japan), 2) WASCA (Zeiss/Meditec, Jena, Germany), and 3) Allegretto (WaveLight Laser Technologies AG, Erlangen, Germany). Subjects were healthy adults (19 men and 21 women; mean age: 20.8 +/- 2.5 years). Refractive errors ranged from +1.5 to -9.75 diopters (D) (mean: +1.83 +/- 2.74 D) with up to 1.75 D cylinder (mean: 0.58 +/- 0.53 D). Three readings were collected per instrument by one examiner without anticholinergic agents. Refraction values were decomposed into vector components for analysis, resulting in mean spherical equivalent refraction (M) and J0 and J45 being vectors of cylindrical power at 0 degrees and 45 degrees, respectively.
RESULTS: Positive correlation was observed between wavefront analyzers and the SRW-5000 for spherical equivalent refraction (OPD-Scan, r=0.959, P<.001; WASCA, r=0.981, P<.001; Allegretto, r=0.942, P<.001). Mean differences and limits of agreement showed more negative spherical equivalent refraction with wavefront analyzers (OPD-Scan, 0.406 +/- 0.768 D [range: 0.235 to 0.580 D] [P<.001]; WASCA, 0.511 +/- 0.550 D [range: 0.390 to 0.634 D] [P<.001]; and Allegretto, 0.434 +/- 0.904 D [range: 0.233 to 0.635 D] [P<.001]). A second analysis eliminating outliers showed the same trend but lower differences: OPD-Scan (n=75), 0.24 +/- 0.41 D (range: 0.15 to 0.34 D) (P<.001); WASCA (n=78), 0.46 +/- 0.47 D (range: 0.36 to 0.57 D) (P<.001); and Allegretto (n=77), 0.30 +/- 0.62 D (range: 0.16 to 0.44 D) (P<.001). No statistically significant differences were noted for J0 and J45.
CONCLUSIONS: Wavefront analyzer refraction resulted in 0.30 D more myopia compared to SRW-5000 refraction in eyes without cycloplegia. This is the result of the accommodation excess attributable to instrument myopia. For the relatively low degrees of astigmatism in this study (<2.0 D), good agreement was noted between wavefront analyzers and the SRW-5000.
Copyright (C) 2006 SLACK Incorporated
KW - accuracy
KW - wavefront analyzers
KW - binocular open-view autorefractor
KW - refractive error
KW - non-cycloplegic eyes
KW - spherical equivalent refraction
KW - myopia
KW - cycloplegia
KW - astigmatism
UR - http://www.scopus.com/inward/record.url?scp=33750117795&partnerID=8YFLogxK
UR - https://www.healio.com/ophthalmology/journals/jrs/2006-10-22-8/%7bc1bde9b9-bf4b-495f-87dc-ff50dcf742f1%7d/wavefront-analyzers-induce-instrument-myopia
U2 - 10.3928/1081-597X-20061001-10
DO - 10.3928/1081-597X-20061001-10
M3 - Article
C2 - 17061717
SN - 1081-597X
VL - 22
SP - 795
EP - 803
JO - Journal of Refractive Surgery
JF - Journal of Refractive Surgery
IS - 8
ER -